Article: Scientists at University of Milan describe research in instrumentation research.(Report)

According to recent research from Milan, Italy, "In this paper, we present a novel procedure for the on-the-field autocalibration of triaxial micro accelerometers, which requires neither any equipment nor a controlled environment and allows increasing the accuracy of this kind of microsensor. The procedure exploits the fact that, in static conditions, the modulus of the accelerometer output vector matches that of the gravity acceleration."

"The calibration model incorporates the bias and scale factor for each axis and the cross-axis symmetrical factors. The parameters are computed through nonlinear optimization, which is solved in a very short time. The ...