Electromagnetic interference produced by a hearing aid device on electrocardiogram recording

Abstract

An 85-year-old male was implanted with a single-chamber permanent pacemaker because of atrial fibrillation with slow ventricular response. The patient had a chronic hearing impairment and decided to buy a hearing aid device. The MyLink device (MyLink, Phonak, Stafa, Switzerland) is a multifrequency FM transmitter/receiver (169.40-176.00 MHz and 214.00-220.00 MHz) with a neckloop antenna that is designed to be used in combination with a second FM transmitter, which detects sound, produced by an audio source or person, and transmits this information to the MyLink wearer. These transmissions are subsequently converted by the MyLink and sent to the patient's existing hearing aids ...